FAQ - Frequently Asked Questions
Error Code:[0x5A63]
Error Message: PDA detector: Flow cell temperature does not reach to set temperature. Change the setting for the cell temperature to be room temperature plus 15ºC min.
- Category :
Answer
«Main Cause»
The temperature did not decrease to the setting temperature even after five minutes of temperature control.
«Countermeasure»
Specify a somewhat high [CELL TEMP] setting in method parameter settings.
Change the cell temperature setting so that it is at least 15ºC higher than room temperature.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.