Error Code:[0x5A54]

Error Message: PDA detector: PDA detector failed to connect to system controller. Check the communication environment.
Category : 

Answer

«Main Cause»
Communication between the instrument and system controller was interrupted.
 
«Countermeasure»
· Check whether a LAN cable is disconnected (PDA detector unit, computer, and hub).
· Check whether the computer power is switched OFF.
· Check whether the hub power is switched OFF.