Error Code:[0x5A45]

Error Message: PDA detector: Intensity is too high because the exposure time of PDAs is too long. Run recalibration on exposure time.
Category : 

Answer

«Main Cause»
Light level exceeded the threshold value due to the long accumulation time.
 
«Countermeasure»
Calibrate the accumulation time.