FAQ - Frequently Asked Questions
Error Code:[0x58CC] ERROR W LAMP NOT CONNECTED
Error Message: UV detector: The connection of W lamp failed to be checked. Check the connection of W lamp. If the error is not resolved, replace W lamp.
- Category :
Answer
«Main Cause»
The W lamp is not connected or the illumination circuit is not functioning properly.
«Countermeasure»
Implement corrective measures in the following sequence.
(1) Check the W lamp connection.
(2) Replace the W lamp.
(1) Check the W lamp connection.
(2) Replace the W lamp.
Click here for Replacement method of Lamp (HPLC)
If that does not resolve the error, switch the instrument power OFF and contact a Shimadzu sales or service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.