FAQ - Frequently Asked Questions
- Products Search > HPLC > Column Oven > Error Code:[0x9401] ERROR GAS DETECTED
Error Code:[0x9401] ERROR GAS DETECTED
Error Message: Column oven: The gas sensor detected organic solvent gas, so the temperature control was stopped. Check leakage of fluid from the flow line, such as the tubing connection of the column.
- Category :
Answer
«Main Cause»
Temperature control was stopped because the gas sensor inside the column oven detected organic solvent leakage.
«Countermeasure»
· If an organic solvent leaks inside the oven, implement corrective measures to prevent tubing leaks.
· Check for organic solvent gases around the instrument. If there are any organic solvent gases, thoroughly ventilate the room.
· If there are no problems with the above two measures, then calibrate the gas sensor sensitivity.
· Check for organic solvent gases around the instrument. If there are any organic solvent gases, thoroughly ventilate the room.
· If there are no problems with the above two measures, then calibrate the gas sensor sensitivity.
Click here for "(LC) Gas Sensor Calibration".