  • No : 4010
  • Date of Issue : 2020/12/02 16:45
  • Print

Error Code:[0x5281] FATAL PCG POWER FAN

Error Message: Plate changer: The temperature control was stopped because a rotation error of the power fan was detected. Restart the instrument.
Category : 

Answer

«Main Cause»
A rotation problem was detected for the fan inside the back of the instrument.
 
«Countermeasure»
Check for foreign matter caught in the fan inside the back of the unit. If that does not resolve the error, switch the power OFF and contact a Shimadzu sales or service representative.