• Font Size
  • S
  • M
  • L
  • No : 3902
  • Displayed Date : 2020/12/02 16:44
  • Print

Error Code:[0x5215] FATAL NEEDLE PROTECTED

Error Message: Autosampler: Excessive force was applied when the needle descended. Check the needle drop position for obstacles.
Category : 

Answer

«Main Cause»
The needle cannot be lowered because an obstruction other than the sample vial or injection port was detected.
 
«Countermeasure»
Switch the instrument power OFF and check inside the instrument for something obstructing the needle. If that does not resolve the error, contact a Shimadzu sales or service representative.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "OKBIZ. for FAQ", which is a product of PKSHA Communication Inc.