How to set the exposure time and how to reset the total lamp usage time for the PDA detector.
- Category :
Answer
1. If the LabSolutions analysis program is running, exit the analysis program.
2. Click the
(Start) menu to display the list of applications and then click [LabSolutions]-[Instrument Maintenance Tools].
3. Select the desired options in the [Tool Launcher] window.
Select [xxx Utility] for [Application]. (For xxx Utility, select the name with the name of the device you want to maintain.)
Select the analysis instrument that is connected to the instrument for maintenance for [Detector Name].
4. Select a registered username from the [User ID] list, enter the password, and click [OK].
5. [xxx Utility] is displayed.
6. Click the
(Wavelength Calibration) button in the [PDA Utility] sub-window.
7. Exposure Time
a) Select [Set Exposure Time] in the [Function] menu.
b) The exposure time is automatically set, and the result is displayed.
c) Check the exposure time value, and click [Yes].
d) The exposure time and the voltage of the tungsten (W) lamp are set.
8. Reset the Total Lamp Usage Time
a) Select [Lamp Time] in the [Function] menu.
b) The total usage time of each lamp is displayed.
c) Clicking [Reset] for each lamp resets the total usage time.