• Font Size
  • S
  • M
  • L
  • No : 3012
  • Date of Issue : 2020/11/09 14:46
  • Print

Selecting a (MR) Mixer

Selecting a (MR) Mixer
Category : 

Answer

Baseline stability is dependent on the flowrate, detector, and mobile phase. Therefore, select a mixer that is appropriate for given system configuration and analytical conditions.
Gradient Type Mixer
High-Pressure Gradient MR 180 μL
Low-Pressure Gradient MR 300 μL LPGE
(To reduce system volume, use a MR 40 μL LPGE mixer.)
 
The table below shows an example of a mixer used to reduce the gradient delay volume in a high-pressure gradient system, such as for ultra high-speed analysis.
It indicates the smallest capacity mixer usable for the given combination of detector and mobile phase.
  Using a Mobile Phase Containing Trifluoroacetic Acid Using a Mobile Phase Containing Formic or Acetic Acids Using a Mobile Phase with Low Absorption, such as Water (Including Buffer Solutions), Organic Solvents, or Mixtures Thereof
Using a Photodiode Array Detector MR 180 μL MR 180 μL MR 180 μL
Using an Absorbance Detector MR 180 μL MR 100 μL MR 40 μL
Using LCMS or LCMSMS - MR 20 μL MR 20 μL
If multiple detectors are connected, use the mixer with the largest capacity. For example, if using LC/MS and PDA detectors, use the MR180 μLII. If the type of mobile phase to be used is unknown, use the largest capacity mixer, which is the MR180 μLII.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.