Cleaning the Needle
Cleaning the needle with a rinse kit.
- Category :
Answer
When you receive an error message that the needle is clogged, it may be clogged with salt or high boiling point components from a salting sample or high temperature analysis. Cleaning with a rinse kit may remove blockage.
Needle Rinse Kit (221-86643-41)
| PN(Parts Number) | Name | Qty |
|---|---|---|
| 228-15672-91 | Adaptor for syringe ![]() |
1 |
| 046-00043-05 | 5mL Syringe ![]() |
1 |
| 228-16004-11 | Coupling
![]() |
1 |
For instructions on use, please refer to the attached instruction manual.