• Font Size
  • S
  • M
  • L

FAQ within [ Supercritical Fluid Chromatograph ]

1-10 of Total 34

1 / 4page
  • Error Code:[0x8B01] SFE SYSTEM ERROR

    «Main Cause» System Error «Countermeasure» Press the (power button) on the instrument to turn the power OFF and contact a Shimadzu sales or service represe... Show Detail

    • No:3685
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B07] SFE ERR DISCONNECT

    «Main Cause» The connection cable between the SFE and the rack-change is not connected properly. Or the rack-changer is turned off. «Countermeasure» Check ... Show Detail

    • No:3686
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B10] SFE ERR NDLE HOME

    «Main Cause» Move to direction was faulty. «Countermeasure» Check if nothing is stuck in the movable part. If nothing is stuck, contact a Shimadzu sales or... Show Detail

    • No:3687
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B11] SFE ERR SLIP

    «Main Cause» SFE: Motor has slipped. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3688
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B12] SFE NDLE PROTECTED

    «Main Cause» Needle tip detects strong force. «Countermeasure» Remedy this problem referring to the following procedure. (1) Replace the needle. (2) Verif... Show Detail

    • No:3689
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B20] SFE ERR EXT-V SENS

    «Main Cause» Could not detect the home position of the SFE extraction valve. «Countermeasure» Press the (power button) on the instrument to turn the power ... Show Detail

    • No:3690
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B21] SFE ERR VNT-V SENS

    «Main Cause» Could not detect the home position of the SFE vent valve. «Countermeasure» Press the (power button) on the instrument to turn the power OFF and... Show Detail

    • No:3691
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B30] SFE ERR LEAK DETECTED

    «Main Cause» Eluent leak can be detected in the SFE. «Countermeasure» Stop the pump flow and then execute any action to stop the eluent leak. See detail In... Show Detail

    • No:3692
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B31] SFE ERR LEAK SENS

    «Main Cause» The leak sensor detection value is abnormal. The leak sensor may be disconnected. «Countermeasure» Press the (power button) on the instrument ... Show Detail

    • No:3693
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8B40] SFE NO VESSEL

    «Main Cause» No extraction vessel was placed in the rack position. «Countermeasure» Set the extraction vessel properry. If the error message is displayed a... Show Detail

    • No:3694
    • Date of Issue:2020/12/02 16:31

1-10 of Total 34

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.