• Font Size
  • S
  • M
  • L

FAQ within [ Solvent Delivery Module ]

21-30 of Total 72

3 / 8page
  • Error Code:[0x9117] FATAL INSIDE FAN

    «Main Cause» A motor cooling fan rotation problem was detected. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service rep... Show Detail

    • No:3878
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x9116] ERROR OPEN DRAIN VALVE

    «Main Cause» Solvent delivery was stopped because pressure became high during purging. «Countermeasure» If an error message is displayed during auto-purging... Show Detail

    • No:3877
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x9115] WARNING MOBILE PHASE D

    «Main Cause» The remaining level of the D mobile phase solution is low. «Countermeasure» Replace the mobile phase and reset the mobile phase quantity. Show Detail

    • No:3876
    • Date of Issue:2020/12/02 16:44
  • Error Code:[0x9114] WARNING MOBILE PHASE C

    «Main Cause» The remaining level of the C mobile phase solution is low. «Countermeasure» Replace the mobile phase and reset the mobile phase quantity. Show Detail

    • No:3875
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x9113] WARNING MOBILE PHASE B

    «Main Cause» The remaining level of the B mobile phase solution is low. «Countermeasure» Replace the mobile phase and reset the mobile phase quantity. Show Detail

    • No:3874
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x9112] WARNING MOBILE PHASE A

    «Main Cause» The remaining level of the A mobile phase solution is low. «Countermeasure» Replace the mobile phase and reset the mobile phase quantity. Show Detail

    • No:3873
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x9111] ERROR LEAK DETECTED

    «Main Cause» Solvent delivery was stopped because leakage was detected in the solvent delivery pump. «Countermeasure» Inspect the tubing connections and oth... Show Detail

    • No:3872
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x9110] ERROR HOME POSITION

    «Main Cause» Solvent delivery is stopped because the home position of a motor cannot be detected. «Countermeasure» Switch the instrument power OFF and then ... Show Detail

    • No:3871
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x910F] ERROR PRESSURE MIN

    «Main Cause» Solvent delivery was stopped because the pressure did not reach the lower limit pressure value. «Countermeasure» Inspect the flow channels for ... Show Detail

    • No:3870
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x910E] ERROR PRESSURE MAX

    «Main Cause» Solvent delivery was stopped because the pressure exceeded the upper pressure limit value. «Countermeasure» Inspect the flow channels for block... Show Detail

    • No:3869
    • Date of Issue:2020/12/02 16:43

21-30 of Total 72

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.