Error Code:[0x58E6] FATAL D2 LAMP NOT LIT
«Main Cause» The D2 lamp failed to illuminate. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of Lamp (HPLC) Show Detail
- No:4082
- Displayed Date:2020/12/02 16:45
- Updated DTM:2020/12/07 11:13
-
Error Code:[0x58E5] ERROR D2 NOT CONNECTED
«Main Cause» D2 lamp is not connected. «Countermeasure» Connect the D2 lamp. Show Detail
- No:4081
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58E4] FATAL D2 LAMP DRIVER
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4080
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58DC] FATAL DETECTOR
«Main Cause» Initialization of internal actions failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representat... Show Detail
- No:4079
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58DA] FATAL DETECTOR
«Main Cause» Initialization of internal actions failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representat... Show Detail
- No:4078
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58D6] FATAL INNER FAN
«Main Cause» A rotation problem was detected for the internal fan. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4077
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58D5] ERROR D2 LAMP NOT LIT
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4076
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58D4] ERROR D2 LAMP NOT LIT
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4075
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58D3] ERROR D2 LAMP NOT LIT
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4074
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x58D1] FATAL FILTER POSITION
«Main Cause» Detection of the filter motor home position failed. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service re... Show Detail
- No:4073
- Displayed Date:2020/12/02 16:45