FAQ within [ Detector ]

61-70 of Total 266

7 / 27page
  • Error Code:[0x5A64]

    «Main Cause» Analysis was started without connecting the flow cell. «Countermeasure» Connect the cell to the main unit. Show Detail

    • No:4102
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A63]

    «Main Cause» The temperature did not decrease to the setting temperature even after five minutes of temperature control. «Countermeasure» Specify a somewhat... Show Detail

    • No:4101
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A61]

    «Main Cause» The temperature-controlled cell control circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shi... Show Detail

    • No:4100
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A57]

    «Main Cause» The analysis start time received from the system controller is later than the current PDA detector time. «Countermeasure» Restart the PDA detec... Show Detail

    • No:4099
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A56]

    «Main Cause» An instrument IP address conflict occurred with another device. «Countermeasure» Resolve the IP address conflict. If that does not resolve the ... Show Detail

    • No:4098
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A55]

    «Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail

    • No:4097
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A54]

    «Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether a LAN cable is disconnected (PDA d... Show Detail

    • No:4096
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A52]

    «Main Cause» Communication between the instrument and workstation was interrupted. «Countermeasure» Check the communication environment and operating status... Show Detail

    • No:4095
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A51]

    «Main Cause» Communication between the instrument and system controller was interrupted. «Countermeasure» · Check whether PDA detector and system controller... Show Detail

    • No:4094
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5A45]

    «Main Cause» Light level exceeded the threshold value due to the long accumulation time. «Countermeasure» Calibrate the accumulation time. Show Detail

    • No:4093
    • Date of Issue:2020/12/02 16:45