• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

51-60 of Total 266

6 / 27page
  • Error Code:[0x5A82]

    «Main Cause» A broken circuit may have occurred in the lamp housing heater. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail

    • No:4112
    • Displayed Date:2020/12/02 16:46
  • Error Code:[0x5A81]

    «Main Cause» A broken circuit may have occurred in the cell heater. «Countermeasure» If the error is not resolved after reconnecting the cell, switch the in... Show Detail

    • No:4111
    • Displayed Date:2020/12/02 16:46
  • Error Code:[0x5A80]

    «Main Cause» A broken circuit may have occurred in the monochromator heater. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales o... Show Detail

    • No:4110
    • Displayed Date:2020/12/02 16:46
  • Error Code:[0x5A7A]

    «Main Cause» A broken circuit may have occurred in the lamp housing sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or... Show Detail

    • No:4109
    • Displayed Date:2020/12/02 16:46
  • Error Code:[0x5A79]

    «Main Cause» A broken circuit may have occurred in the leak sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service... Show Detail

    • No:4108
    • Displayed Date:2020/12/02 16:46
  • Error Code:[0x5A78]

    «Main Cause» A leak sensor short circuit may have occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service represe... Show Detail

    • No:4107
    • Displayed Date:2020/12/02 16:45
  • Error Code:[0x5A70]

    «Main Cause» A broken circuit may have occurred in the cell temperature sensor. «Countermeasure» If the error is not resolved after reconnecting the cell, s... Show Detail

    • No:4106
    • Displayed Date:2020/12/02 16:45
  • Error Code:[0x5A69]

    «Main Cause» Cannot write data to D2 memory. «Countermeasure» Check the D2 lamp connection. If that does not resolved the error, replace the D2 ... Show Detail

    • No:4105
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:14
  • Error Code:[0x5A68]

    «Main Cause» Information in the D2 lamp memory could not be read. «Countermeasure» Contact a Shimadzu sales or service representative. Show Detail

    • No:4104
    • Displayed Date:2020/12/02 16:45
  • Error Code:[0x5A65]

    «Main Cause» Analysis was started without registering the cell in the main unit. «Countermeasure» Register the cell in the main unit using Utility. Show Detail

    • No:4103
    • Displayed Date:2020/12/02 16:45

51-60 of Total 266

This site uses the FAQ system "OKBIZ. for FAQ", which is a product of PKSHA Communication Inc.