• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

51-60 of Total 266

6 / 27page
  • Error Code:[0x8922]

    «Main Cause» Error occurred during system check. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3645
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8941]

    «Main Cause» There is a PC data buffer overflow. «Countermeasure» Check the PC status and the cable connections. If the error persists, turn the power OFF ... Show Detail

    • No:3646
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8942]

    «Main Cause» Hard disk initialization failed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3647
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8943]

    «Main Cause» Hard disk is full. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3648
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8944]

    «Main Cause» Detector buffer memory overflowed. «Countermeasure» Turn the power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3649
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8945]

    «Main Cause» The signal is saturated because the exposure time for the PDA elements is inappropriate. «Countermeasure» Perform exposure time optimization u... Show Detail

    • No:3650
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8952]

    «Main Cause» Ethernet communication with the PC timed out. «Countermeasure» Check the power supply of the instrument and cable connections. Restart the PDA... Show Detail

    • No:3651
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8954]

    «Main Cause» Ethernet communication with the CBM timed out. «Countermeasure» Check the network environment. Turn the power of PDA detector and CBM OFF once... Show Detail

    • No:3652
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8956]

    «Main Cause» A device with the same IP address as the instrument exists on the same network. «Countermeasure» Change the IP address of the instrument or de... Show Detail

    • No:3653
    • Date of Issue:2020/12/02 16:31
  • Error Code:[0x8961]

    «Main Cause» Temperature-controlled flow cell malfunctioned, control circuit malfunctioned. «Countermeasure» Turn the power OFF and contact a Shimadzu sale... Show Detail

    • No:3654
    • Date of Issue:2020/12/02 16:31

51-60 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.