FAQ within [ Detector ]
-
«Main Cause» The lamp is turned off because the temperature sensor of the detector lamp housing is malfunctioning. «Countermeasure» Press the (power switc... Show Detail
- No:3416
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail
- No:3417
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:15
-
«Main Cause» A lighting defect has occurred due to a broken filament in the D2 lamp or a fault in the constant current circuit, a lighting malfunction has... Show Detail
- No:3418
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:15
-
«Main Cause» This is a lighting defect caused by a broken filament in the W lamp, or due to a fault in the lighting circuit. «Countermeasure» Re... Show Detail
- No:3419
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:16
-
«Main Cause» The lamp house cover is open or there is a fault in the cover sensor. «Countermeasure» Close the lamp house cover. If the error is not resolve... Show Detail
- No:3420
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail
- No:3421
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:16
-
«Main Cause» The lamp is turned off because the temperature of the detector lamp housing is high. «Countermeasure» Follow the procedures below in the foll... Show Detail
- No:3422
- Date of Issue:2020/12/02 16:10
-
«Main Cause» There is a fault in the W lamp lighting circuit. «Countermeasure» Replace the W lamp. Click here for Replacement method of Lamp ... Show Detail
- No:3423
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:17
-
«Main Cause» There is a fault in the D2 lamp or a fault in the lighting circuit. «Countermeasure» Replace the D2 lamp. Click here for Replace... Show Detail
- No:3424
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:17
-
«Main Cause» The D2 lamp connector is not connected. Alternatively, the D2 lamp cable is broken or there is a fault in the board. «Countermeasure»... Show Detail
- No:3425
- Date of Issue:2020/12/02 16:10
- Last Updated:2020/12/07 11:18