• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

261-266 of Total 266

27 / 27page
  • Maintenance for SPD-40(V) (Video)

    Show Detail

    • No:4301
    • Date of Issue:2020/12/03 11:16
    • Last Updated:2020/12/04 14:53
    • Brached FAQ
  • Maintenance for SPD-M40 (Video)

    Part No. 228-53924-05 Dust Filter 1. Turn OFF power to the instrument. 2. Open the filter cover (A) and then remove the filter (B). ... Show Detail

    • No:4309
    • Date of Issue:2020/12/03 13:04
    • Last Updated:2020/12/04 14:53
    • Brached FAQ
  • Replacement of Lamp (HPLC)

    If you need a video, click here forSPD-M40 maintenance. WARNING: •Before replacing the lamp, turn the instrument power switch OFF and unplug it. ... Show Detail

    • No:4424
    • Date of Issue:2020/12/04 16:01
    • Last Updated:2021/07/12 14:11
    • Brached FAQ
  • (HPLC) RID "OVER" is displayed

    • "OVER" is displayed in the "range" field of the initial screen. Cause: The recorder output has exceeded the scale limit. Actions: Press [zero]. ... Show Detail

    • No:4750
    • Date of Issue:2021/01/12 13:50
    • Last Updated:2021/01/12 13:51
  • (HPLC) RID "EN.ALRM" is displayed

    • Causes: The incident light intensity to the photodiode is insufficient. * The total light intensity (TOTAL EN) is less than 5500.(In case of RID-10A, les... Show Detail

    • No:4753
    • Date of Issue:2021/01/12 13:56
  • (HPLC) SPD-M30A Handling and Cleaning of cell

    <Cell Cleaning Method> 1. Flow 17 % phosphoric acid (5times dilution of 85 % phosphoric acid in water) for one hour at 1 mL/min. 2. Switch to water a... Show Detail

    • No:4755
    • Date of Issue:2021/01/12 14:08

261-266 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.