• Font Size
  • S
  • M
  • L

FAQ within [ Detector ]

111-120 of Total 266

12 / 27page
  • Error Code:[0x5803] WARNING MICRO BOARD

    «Main Cause» The micro cell DET board is not properly recognized. «Countermeasure» Check the cell type. To replace the current DET board with one for a micr... Show Detail

    • No:4052
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x58E9] WARNING NOT TEMP-CONT CELL

    «Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail

    • No:4051
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x80F7]

    «Main Cause» The PDA detector independently started up the system controller during an analysis. «Countermeasure» Either stop the PDA detector analysis or w... Show Detail

    • No:3813
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F5]

    «Main Cause» The PDA detector system name was changed during use. «Countermeasure» The parent-child relationship might have failed between the system contro... Show Detail

    • No:3812
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F4]

    «Main Cause» The system name setting for the master PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed be... Show Detail

    • No:3811
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F3]

    «Main Cause» The IP address for the PDA detector was changed during use. «Countermeasure» The parent-child relationship might have failed between the system... Show Detail

    • No:3810
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80F0]

    «Main Cause» Multiple PDA detectors attempted to connect to the same system controller (redundant parent-child relationships). «Countermeasure» Change PDA d... Show Detail

    • No:3809
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x80E0]

    «Main Cause» Eleven or more system controllers and PDA detectors are assigned to the same group. «Countermeasure» Specify ten or fewer devices per group. Show Detail

    • No:3808
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x8084]

    «Main Cause» A separate workstation than connected to the system controller is connected to the PDA detector. (System controllers and PDA detectors with an e... Show Detail

    • No:3807
    • Date of Issue:2020/12/02 16:43
  • Error Code:[0x2F28]

    «Main Cause» PDA detector: PDA detector failed to connect to the workstation. «Countermeasure» Check the communication environment (LAN cables, hub, etc.). ... Show Detail

    • No:3792
    • Date of Issue:2020/12/02 16:43

111-120 of Total 266

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.