FAQ within [ Detector ]

91-100 of Total 266

10 / 27page
  • Error Code:[0x58CD] FATAL W LAMP NOT LIT

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures in the following sequence. (1... Show Detail

    • No:4072
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:13
  • Error Code:[0x58CC] ERROR W LAMP NOT ...

    «Main Cause» The W lamp is not connected or the illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures... Show Detail

    • No:4071
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:12
  • Error Code:[0x58CA] ERROR D2 LAMP NOT...

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures in the following se... Show Detail

    • No:4070
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:11
  • Error Code:[0x58C9] ERROR W LAMP NOT LIT

    «Main Cause» The W lamp or W lamp illumination circuit is not functioning properly. «Countermeasure» Implement corrective measures in the follow... Show Detail

    • No:4069
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:11
  • Error Code:[0x58C7] ERROR D2 LAMP NOT...

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4068
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:10
  • Error Code:[0x58C6] WORNING LAMP COVE...

    «Main Cause» The lamp housing opened or the cover sensor is not functioning properly. «Countermeasure» Close the lamp housing cover. If that does not resolv... Show Detail

    • No:4067
    • Displayed Date:2020/12/02 16:45
  • Error Code:[0x58C4] ERROR W LAMP NOT LIT

    «Main Cause» The W lamp failed to illuminate due to a broken filament or the illumination circuit is not functioning properly. «Countermeasure» ... Show Detail

    • No:4066
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:10
  • Error Code:[0x58C3] ERROR D2 LAMP NOT...

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4065
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:09
  • Error Code:[0x58C2] ERROR D2 LAMP NOT...

    «Main Cause» The D2 lamp or illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacemen... Show Detail

    • No:4064
    • Displayed Date:2020/12/02 16:45
    • Updated DTM:2020/12/07 11:08
  • Error Code:[0x5885] FATAL REAR FAN

    «Main Cause» A rotation problem was detected for the fan on the back of the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shima... Show Detail

    • No:4063
    • Displayed Date:2020/12/02 16:45

91-100 of Total 266

This site uses the FAQ system "OKBIZ. for FAQ", which is a product of PKSHA Communication Inc.