FAQ - Frequently Asked Questions
- Products Search > HPLC > Common
Keyword Search
Frequently Referenced FAQ
FAQ within [ Common ]
-
The baseline undulates with the pump stroke
Click [PURGE] to purge the bubbles. Alternatively, connect a syringe to the drain tube connection port to suction out the bubbles. Click here for Manual P... Show Detail
- No:4296
- Date of Issue:2020/12/03 11:00
- Brached FAQ
-
Check for leakage from cell inlet and outlet pipe connections and from the cell window. If there is leakage, disassemble and clean the cell. If there is lo... Show Detail
- No:4295
- Date of Issue:2020/12/03 10:59
- Brached FAQ
-
Inspect the suction filter for blockage. Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowr... Show Detail
- No:4294
- Date of Issue:2020/12/03 10:57
- Brached FAQ
-
Inspect the suction filter for blockage.Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowra... Show Detail
- No:4293
- Date of Issue:2020/12/03 10:52
- Brached FAQ
-
Check the column using known analytical conditions. If the column has deteriorated, replace it. Confirm whether the solvent delivery pump is functioning pr... Show Detail
- No:4292
- Date of Issue:2020/12/03 10:49
- Brached FAQ
-
Check the column using known analytical conditions. If the column has deteriorated, replace it. Connect the tubing correctly. Backflush the high pressure... Show Detail
- No:4291
- Date of Issue:2020/12/03 10:46
- Brached FAQ
-
Pump 2-propanol to rinse the check valve. However, if a buffer solution is being used, rinse the check valve with purified water first. Then rinse it by pump... Show Detail
- No:4290
- Date of Issue:2020/12/03 10:43
- Brached FAQ
-
No power is supplied when power switch is switched ON
Plug in the power cord properly. Replace it with a power cord of the same type. Use a power supply that meets the power supply specifications for the ins... Show Detail
- No:4279
- Date of Issue:2020/12/03 10:18
- Brached FAQ
-
Autosamplers include rinsing functionality. Rinse the exterior and interior surfaces of the needle with solvent. Change the type of rinse solution and rins... Show Detail
- No:4278
- Date of Issue:2020/12/03 10:16
- Brached FAQ
-
Poor Area Value Reproducibility
If the solvent delivery pump flow is unstable, then area values and reproducibility are also unstable. Check the solvent delivery pump for pulsation, leakage... Show Detail
- No:4276
- Date of Issue:2020/12/03 10:09
- Brached FAQ