• Font Size
  • S
  • M
  • L

FAQ within [ Column Oven ]

41-50 of Total 76

5 / 8page
  • Error Code:[0x8636] ERR POST COOLER

    «Main Cause» The post-column cooler is not rotating correctly. «Countermeasure» If anything is interfering with the fan rotation, turn the power OFF and rem... Show Detail

    • No:3628
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8632]

    «Main Cause» Cannot write to oven :Management Device (CMD). «Countermeasure» Once turn off the power, and turn on the power again. If this message still app... Show Detail

    • No:3627
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x8630] ERR NO RV-* HOME

    «Main Cause» Rotation error in an automatic column-switching valve. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3626
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0079] COOLER NOT EXIST

    «Main Cause» Cooler is not mounted or the cable is snapping. «Countermeasure» Turn power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:3581
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0077] WARNING: LOW SET TEMP

    «Main Cause» For CTO-10AC (vp) or CTO-20AC: If the COOLER MODE is "1" or "2", the room temperature is less than -10ºC. When "0" is set, Settings of oven temp... Show Detail

    • No:3580
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0076] ERR LEAK SENS

    «Main Cause» Leak sensor malfunctions (probably due to broken/disconnected wire). «Countermeasure» Turn power OFF and contact a Shimadzu sales or service re... Show Detail

    • No:3579
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0074] ERR FAN

    «Main Cause» Fan speed does not increase to the set value. «Countermeasure» If the cause is obstruction of the fan by tubing or other object, turn the power... Show Detail

    • No:3578
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0073] ERR LEAK DETECT

    «Main Cause» Density of solvent vapor inside column oven has exceeded leak sensor actuation level. «Countermeasure» (1) Mobile phase has leaked inside the o... Show Detail

    • No:3577
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0072] ERR SENSOR L

    «Main Cause» Thermosensor readings are abnormally low. The sensor may be disconnected. «Countermeasure» Turn power OFF and contact a Shimadzu sales or servi... Show Detail

    • No:3576
    • Date of Issue:2020/12/02 16:30
  • Error Code:[0x0071] ERR SENSOR H

    «Main Cause» Thermosensor readings are abnormally high. There may be a short circuit in the sensor. «Countermeasure» Turn power OFF and contact a Shimadzu s... Show Detail

    • No:3575
    • Date of Issue:2020/12/02 16:30

41-50 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.