• Font Size
  • S
  • M
  • L

FAQ within [ Column Oven ]

21-30 of Total 76

3 / 8page
  • Error Code:[0x9401] ERROR GAS DETECTED

    «Main Cause» Temperature control was stopped because the gas sensor inside the column oven detected organic solvent leakage. «Countermeasure» · ... Show Detail

    • No:4037
    • Date of Issue:2020/12/02 16:45
    • Last Updated:2022/05/09 15:32
  • Error Code:[0x8635] FATAL REAR FAN

    «Main Cause» A fan rotation problem was detected in the power supply heat exhaust fan. «Countermeasure» Switch the instrument power OFF and remove any forei... Show Detail

    • No:4036
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x8634] FATAL AC HEATER

    «Main Cause» The heater is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4035
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x8633] FATAL GAS SENSOR

    «Main Cause» A broken circuit may have occurred in the gas sensor. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:4034
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x540B] ERROR INT-VALVE2 [0x##]

    «Main Cause» An error was detected in the flow channel selection valve 2 inside the column oven. (The error code "0x##" indicates a valve unit error. Refer t... Show Detail

    • No:4033
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x540A] ERROR LEAK DETECTED

    «Main Cause» Temperature control was stopped because leakage was detected in the column oven. «Countermeasure» Inspect the leakage from flow channels or oth... Show Detail

    • No:4032
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5409] FATAL ROOM TEMP SENSOR

    «Main Cause» A broken or short circuit may have occurred in the room temperature sensor. «Countermeasure» Switch the instrument power OFF and contact a Shim... Show Detail

    • No:4031
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5408] ERROR CMD CHECK SUM

    «Main Cause» Data written in the column management device (CMD) is inconsistent. «Countermeasure» Disconnect or reinitialize the CMD. Show Detail

    • No:4030
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5407] FATAL RAM FAILURE

    «Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4029
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5406] FATAL ROM FAILURE

    «Main Cause» A system error occurred. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service representative. Show Detail

    • No:4028
    • Date of Issue:2020/12/02 16:45

21-30 of Total 76

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.