Error Code:[0x5274] ERROR PCG PLATE SET
«Main Cause» A separate sample plate was placed in the stack where the plate in the autosampler was previously placed. «Countermeasure» Check the sample plate. Show Detail
- No:3997
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x5273] ERROR PCG PLATE BACK
«Main Cause» The arm chuck cannot release the sample plate. «Countermeasure» Restart the instrument. Show Detail
- No:3996
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x5272] ERROR PCG CHUCK CLOSE
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3995
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x5271] ERROR PCG CHUCK OPEN
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3994
- Displayed Date:2020/12/02 16:45
-
Error Code:[0x5270] ERROR PCG CHUCK HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3993
- Displayed Date:2020/12/02 16:44
-
Error Code:[0x526F] ERROR PCG Y SLIP
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3992
- Displayed Date:2020/12/02 16:44
-
Error Code:[0x526E] ERROR PCG Y HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3991
- Displayed Date:2020/12/02 16:44
-
Error Code:[0x526D] ERROR PCG Z SLIP
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3990
- Displayed Date:2020/12/02 16:44
-
Error Code:[0x526C] ERROR PCG Z HOME
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3989
- Displayed Date:2020/12/02 16:44
-
Error Code:[0x526B] ERROR PCG THETA SLIP
«Main Cause» Something is obstructing the arm's line of movement or there is mechanical problem in the drive unit. «Countermeasure» Check for any obstructio... Show Detail
- No:3988
- Displayed Date:2020/12/02 16:44
61-70 of Total 297