  • Error Code:[0x5291] ERROR ACQUISITION TIME

    «Main Cause» Autosampler: Repeated injections did not complete within the data acquisition time. «Countermeasure» Set the injection interval to a time longe... Show Detail

    • No:4151
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x5290] FATAL SYRINGE DATA

    «Main Cause» Autosampler: The data stored in the syringe unit is not correct, so the data was initialized. «Countermeasure» Although this does not affect th... Show Detail

    • No:4150
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528F] ERROR SYRINGE LEAK

    «Main Cause» Autosampler: The syringe unit detected leak. «Countermeasure» Wipe off the leaking liquid and check the leaking part. If fluid is leaking from ... Show Detail

    • No:4149
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528E] FATAL SYRINGE LEAK SEN

    «Main Cause» Autosampler: There is an error in the leak sensor of the syringe unit. The leakage sensor may be broken. «Countermeasure» Turn off the instrume... Show Detail

    • No:4148
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528D] FATAL SYRINGE LPV HOME

    «Main Cause» Autosampler: The 3 way valve of the syringe unit failed to detect the home position. «Countermeasure» Turn off the instrument and contact a Shi... Show Detail

    • No:4147
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x528C] FATAL SYRINGE HOME

    «Main Cause» Autosampler: The syringe unit failed to detect the home position. «Countermeasure» Check the syringe plunger and plunger holder for foreign obj... Show Detail

    • No:4146
    • Date of Issue:2020/12/02 16:46
  • Error Code:[0x81E5] FATAL LEAK SENSOR

    «Main Cause» The leak sensor cannot be detected properly. The leak sensor might not be connected. «Countermeasure» Switch the instrument power OFF and conta... Show Detail

    • No:4021
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x528B] WARNING SMP, L: -

    «Main Cause» An error occurred in the sample table being executed. The row number of the row with an error is indicated to the right of "L:" and the error ty... Show Detail

    • No:4020
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x528A] ERROR HPV2 ERROR

    «Main Cause» An error was detected from the HPV2 unit connected to the autosampler. «Countermeasure» Check the HPV2 unit. If that does not resolve the error... Show Detail

    • No:4019
    • Date of Issue:2020/12/02 16:45
  • Error Code:[0x5289] ERROR FCV8 ERROR

    «Main Cause» An error was detected from the FCV8 unit connected to the autosampler. «Countermeasure» Check the FCV8 unit. If that does not resolve the error... Show Detail

    • No:4018
    • Date of Issue:2020/12/02 16:45