• Font Size
  • S
  • M
  • L

FAQ within [ Autosampler ]

271-280 of Total 297

28 / 30page
  • Error Code:[0x9232]

    «Main Cause» The autosampler needle Y-direction motor has slipped. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sa... Show Detail

    • No:3463
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9231]

    «Main Cause» The autosampler needle X-direction motor has slipped. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sa... Show Detail

    • No:3462
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9230]

    «Main Cause» The sample vial number that is outside of the setting range for the set plate is specified in the sequence. «Countermeasure» Check the sequence... Show Detail

    • No:3461
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x922F]

    «Main Cause» The left side sample cooler temperature adjustment was stopped because the right side rack was not set for more than 30 minutes. «Countermeasur... Show Detail

    • No:3460
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x922E]

    «Main Cause» The left side sample cooler temperature adjustment was stopped because the left side rack was not set for more than 30 minutes. «Count... Show Detail

    • No:3459
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2022/01/06 09:43
  • Error Code:[0x922D]

    «Main Cause» Racks that have the same ID are set in the autosampler. «Countermeasure» Change the rack ID of either the left or right side rack. Show Detail

    • No:3458
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x922C]

    «Main Cause» The rack ID is not correct. «Countermeasure» Set the correct rack ID for the rack. Show Detail

    • No:3457
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x922B]

    «Main Cause» The position of the needle in relation to the right side rack has not been adjusted. «Countermeasure» Adjust the position of the ne... Show Detail

    • No:3456
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2022/01/06 09:42
  • Error Code:[0x922A]

    «Main Cause» The position of the needle in relation to the left side rack has not been adjusted. «Countermeasure» Adjust the position of the nee... Show Detail

    • No:3455
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2022/01/06 09:39
  • Error Code:[0x9229]

    «Main Cause» Acquisition stopped because the position of the needle in relation to the rack has not been adjusted. «Countermeasure» Adjust the position of t... Show Detail

    • No:3454
    • Date of Issue:2020/12/02 16:10

271-280 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.