• Font Size
  • S
  • M
  • L

FAQ within [ Autosampler ]

251-260 of Total 297

26 / 30page
  • Error Code:[0x9215]

    «Main Cause» Temperature control is turned off because the temperature sensor in the autosampler dehumidifier unit is malfunctioning. «Countermeasure» Press... Show Detail

    • No:3483
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9214]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack R has detected an abnormal condition. «Countermeasure» Pre... Show Detail

    • No:3482
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9213]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack R is malfunctioning. «Countermeasure» Press the (power swi... Show Detail

    • No:3481
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x9210]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack L has detected an abnormal condition. «Countermeasure» Pre... Show Detail

    • No:3480
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920F]

    «Main Cause» Temperature control is turned off because the temperature sensor in autosampler rack L is malfunctioning. «Countermeasure» Press the (power swi... Show Detail

    • No:3479
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920E]

    «Main Cause» The leak sensor in the autosampler high pressure valve is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and c... Show Detail

    • No:3478
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920D]

    «Main Cause» The leak sensor in the autosampler high pressure valve has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn th... Show Detail

    • No:3477
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920C]

    «Main Cause» The leak sensor in the autosampler low pressure valve is malfunctioning. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

    • No:3476
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920B]

    «Main Cause» The leak sensor in the autosampler low pressure valve has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the... Show Detail

    • No:3475
    • Date of Issue:2020/12/02 16:10
  • Error Code:[0x920A]

    «Main Cause» Temperature control is turned off because the autosampler rack R heater has overheated and the thermal fuse has blown. «Countermeasure» Press t... Show Detail

    • No:3474
    • Date of Issue:2020/12/02 16:10

251-260 of Total 297

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.