  • Error Code:[0x9363]

    «Main Cause» The temperature control cell temperature does not lower to the set temperature even after 5 minutes have elapsed. «Countermeasure» Set the [Cel... Show Detail

    • No:3412
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9361]

    «Main Cause» There is a fault in the control circuit of the temperature control cell. «Countermeasure» Replace the temperature control cell. Show Detail

    • No:3411
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9356]

    «Main Cause» The IP address of the instrument is duplicated on a different device. «Countermeasure» Change the IP addresses so there are no duplicates. If ... Show Detail

    • No:3410
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x9353]

    «Main Cause» The detector cell temperature does not reach the set temperature. «Countermeasure» Replace the temperature control cell. If the error is not ... Show Detail

    • No:3409
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9352]

    «Main Cause» Communication between the instrument and the workstation was interrupted. «Countermeasure» Check the communication environment and operating st... Show Detail

    • No:3408
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9345]

    «Main Cause» Light intensity was exceeded because the detector took a long exposure time. «Countermeasure» Calibrate the exposure time. Show Detail

    • No:3407
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9344]

    «Main Cause» Recovery failed when resending transmission data from the instrument to the workstation «Countermeasure» Check the communication environment a... Show Detail

    • No:3406
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9341]

    «Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the communication environm... Show Detail

    • No:3405
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9321]

    «Main Cause» A liquid leak has occurred in the detector. «Countermeasure» Inspect the detector for liquid leaks, and wipe away any liquids from around the l... Show Detail

    • No:3404
    • Date of Issue:2020/12/02 16:09
    • Category: Detector
  • Error Code:[0x9304]

    «Main Cause» There is a fault in the non-volatile memory of the detector or a non-volatile memory write error has occurred. «Countermeasure» Perform calibra... Show Detail

    • No:3403
    • Date of Issue:2020/12/02 16:09
    • Category: Detector