FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
«Main Cause» Could not detect the grating home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shim... Show Detail
- No:3442
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» Could not detect the filter home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shima... Show Detail
- No:3441
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» Failed to receive the detector signal. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service ... Show Detail
- No:3440
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» Failed to receive the detector signal. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service ... Show Detail
- No:3439
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» The heater of the spectrmeter is short or there is a fault in the circuit. «Countermeasure» Press the (power switch) to turn the power OFF and ... Show Detail
- No:3438
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» The optical unit temperature does not reach the set temperature. Temperature control is turned off. «Countermeasure» Contact a Shimadzu sales o... Show Detail
- No:3437
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» The instrument is turned off because the ADC module on the board is malfunctioning. The instrument is stopped. «Countermeasure» Press the (pow... Show Detail
- No:3436
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» Temperature control is turned off because the temperature sensor of the optical unit is malfunctioning. «Countermeasure» Press the (power switc... Show Detail
- No:3435
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» Temperature control is turned off because the temperature of the optical unit is high. «Countermeasure» Press the (power switch) to turn the po... Show Detail
- No:3434
- Date of Issue:2020/12/02 16:10
- Category: Detector
-
«Main Cause» The lamp is turned off because a rotation error was detected in the detector fan. «Countermeasure» Press the (power switch) to turn the power O... Show Detail
- No:3433
- Date of Issue:2020/12/02 16:10
- Category: Detector
841-850 of Total 929