Maintenance for SIL-40 series (Video)
Consumable Parts Number:228-45707-91 SUS Filter Element « Below are the changes for SIL-40C XSi» • Ring(B) is added to S... Show Detail
- No:4320
- Displayed Date:2020/12/03 13:40
- Category: Autosampler
- Brached FAQ
-
Maintenance for SPD-M40 (Video)
Part No. 228-53924-05 Dust Filter 1. Turn OFF power to the instrument. 2. Open the filter cover (A) and then remove the filter (B). ... Show Detail
- No:4309
- Displayed Date:2020/12/03 13:04
- Updated DTM:2020/12/04 14:53
- Category: Detector
- Brached FAQ
-
Maintenance for SPD-40(V) (Video)
- No:4301
- Displayed Date:2020/12/03 11:16
- Updated DTM:2020/12/04 14:53
- Category: Detector
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4300
- Displayed Date:2020/12/03 11:06
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4299
- Displayed Date:2020/12/03 11:05
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Check the light level. If the light level is low even with an air cell or if the lamp has been operated for more than 2000 hours, then replace the lamp. ... Show Detail
- No:4298
- Displayed Date:2020/12/03 11:03
- Category: Detector
- Brached FAQ
-
Specify an appropriate range value. Set [LAMP] to [D2]. The light source lamp illuminates. Click the [ZERO] button in the [RUN] window. Adjust the zero p... Show Detail
- No:4297
- Displayed Date:2020/12/03 11:01
- Category: Detector
- Brached FAQ
-
The baseline undulates with the pump stroke
Click [PURGE] to purge the bubbles. Alternatively, connect a syringe to the drain tube connection port to suction out the bubbles. Click here for Manual P... Show Detail
- No:4296
- Displayed Date:2020/12/03 11:00
- Category: Common , Solvent Delivery Module
- Brached FAQ
-
Check for leakage from cell inlet and outlet pipe connections and from the cell window. If there is leakage, disassemble and clean the cell. If there is lo... Show Detail
- No:4295
- Displayed Date:2020/12/03 10:59
- Category: Common
- Brached FAQ
-
Inspect the suction filter for blockage. Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowr... Show Detail
- No:4294
- Displayed Date:2020/12/03 10:57
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ