FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Drift does not stabilize (Ion Chromatograph)
Hydrate the suppressor and use it as usual. Delivering eluent without hydrating the suppressor may cause sodium ion to enter a section from which the ion is... Show Detail
- No:4387
- Date of Issue:2020/10/26 14:24
- Category: Ion Chromatograph
-
Noises or fluctuation became large (Ion Chromatograph)
Bubbles have accumulated. Pass water through the suppressor. When a large noise persists even after taking the action above, perform "Regeneration Procedure... Show Detail
- No:4386
- Date of Issue:2020/10/26 14:27
- Category: Ion Chromatograph
-
If an error message fails to appear when the mobile phase leaks, or if an error message appears, althoughthere is no leakage, the leak sensor sensitivity mus... Show Detail
- No:8576
- Date of Issue:2020/10/30 11:19
- Category: Column Oven
- Brached FAQ
-
Manual Purging the Solvent Delivery Pump
If adequate purging cannot be achieved with auto-purging or a suction tube does not contain liquid, for example, use the following procedure to manually purg... Show Detail
- No:3011
- Date of Issue:2020/11/09 14:29
- Last Updated:2024/04/11 16:05
- Category: Solvent Delivery Module
-
Baseline stability is dependent on the flowrate, detector, and mobile phase. Therefore, select a mixer that is appropriate for given system configuration and... Show Detail
- No:3012
- Date of Issue:2020/11/09 14:46
- Category: Common
-
Setting the compressibility of solvent (COMP)
Although the instrument corrects the compressibility to suppress pulse increase caused by the effect of the compressibility of the solvent, setting a compres... Show Detail
- No:3013
- Date of Issue:2020/11/09 15:05
- Category: Solvent Delivery Module
-
How to set the exposure time and how to reset the total l...
1. If the LabSolutions analysis program is running, exit the analysis program. 2. Click the (Start) menu to display the list of applications and then cli... Show Detail
- No:3015
- Date of Issue:2020/11/09 15:36
- Category: Detector
-
Rinse the instrument flow channels with purified or distilled water after pumping a buffer solution. If no analyses will be performed for a while, rinse and... Show Detail
- No:3023
- Date of Issue:2020/11/10 14:22
- Category: Common
-
LAN communication confirmation method
You can check the LAN communication between the PC and the instruments by the following method. 1. Right-click on the (Start) menu and select [Run...]... Show Detail
- No:3024
- Date of Issue:2020/11/10 14:41
- Category: Common
-
Cleaning/Replacing the ELSD Nebulizer
CAUTION: Take sufficient care while handling the nebulizer • Never touch the IN tubing fitting section (B) of the nebulizer. • Replace the nebulizer if it ... Show Detail
- No:3072
- Date of Issue:2020/11/11 09:50
- Last Updated:2023/10/18 10:52
- Category: Detector
21-30 of Total 929