Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Inspect the suction filter for blockage. Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowr... Show Detail
- No:4294
- Date of Issue:2020/12/03 10:57
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Inspect the suction filter for blockage.Also rinse them as described below. 1) Flush with 17 % aqueous phosphate solution for about 30 minutes at a flowra... Show Detail
- No:4293
- Date of Issue:2020/12/03 10:52
- Category: Common , Autosampler , Solvent Delivery Module
- Brached FAQ
-
Check the column using known analytical conditions. If the column has deteriorated, replace it. Confirm whether the solvent delivery pump is functioning pr... Show Detail
- No:4292
- Date of Issue:2020/12/03 10:49
- Category: Common , Autosampler , Column Oven , Detector , Solvent Delivery Module
- Brached FAQ
-
Check the column using known analytical conditions. If the column has deteriorated, replace it. Connect the tubing correctly. Backflush the high pressure... Show Detail
- No:4291
- Date of Issue:2020/12/03 10:46
- Category: Common , Autosampler , Column Oven
- Brached FAQ
-
Pump 2-propanol to rinse the check valve. However, if a buffer solution is being used, rinse the check valve with purified water first. Then rinse it by pump... Show Detail
- No:4290
- Date of Issue:2020/12/03 10:43
- Category: Common , Autosampler , Column Oven , Solvent Delivery Module
- Brached FAQ
-
The pressure change is large when high pressure valve swi...
Disassemble and clean the high pressure valve. If disassembled cleaning is not effective, then replace the rotor and stator. Contact a Shimadzu sales or se... Show Detail
- No:4289
- Date of Issue:2020/12/03 10:38
- Category: Autosampler
- Brached FAQ
-
Confirm whether the baseline changes significantly when the cell outlet pipe is blocked. Alternatively, visually check for any bubbles flowing in the flow ce... Show Detail
- No:4288
- Date of Issue:2020/12/03 10:36
- Category: Detector
- Brached FAQ
-
Cell Temperature Control Failure
Make sure the connector is connected securely. Specify an appropriate cell temperature-control temperature. The cell temperature-control setting range vari... Show Detail
- No:4287
- Date of Issue:2020/12/03 10:34
- Category: Detector
- Brached FAQ
-
Close column oven door. Specify an appropriate column temperature-control temperature. The possible temperature-control range varies depending on the room ... Show Detail
- No:4286
- Date of Issue:2020/12/03 10:33
- Category: Column Oven
- Brached FAQ
-
The rinse solution level for the seal automatic rinsing k...
Inspect (replace) the diaphragm. Tighten the connection. If the leak persists after tightening the connection, then replace the connection. Show Detail
- No:4285
- Date of Issue:2020/12/03 10:31
- Category: Solvent Delivery Module
- Brached FAQ