FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Consumable Part No. 228-72570-01 Air filter 1. Remove the filter holder (A) on the right side by pressing its right and left knobs (a). 2. ... Show Detail
- No:4339
- Date of Issue:2020/12/03 13:45
- Category: Column Oven
- Brached FAQ
-
Maintenance for SIL-40 series (Video)
Consumable Parts Number:228-45707-91 SUS Filter Element « Below are the changes for SIL-40C XSi» • Ring(B) is added to S... Show Detail
- No:4320
- Date of Issue:2020/12/03 13:40
- Category: Autosampler
- Brached FAQ
-
Maintenance for SPD-M40 (Video)
Part No. 228-53924-05 Dust Filter 1. Turn OFF power to the instrument. 2. Open the filter cover (A) and then remove the filter (B). ... Show Detail
- No:4309
- Date of Issue:2020/12/03 13:04
- Last Updated:2020/12/04 14:53
- Category: Detector
- Brached FAQ
-
Maintenance for SPD-40(V) (Video)
- No:4301
- Date of Issue:2020/12/03 11:16
- Last Updated:2020/12/04 14:53
- Category: Detector
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4300
- Date of Issue:2020/12/03 11:06
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Check whether the noise disappears when solvent delivery stops. Check the solvent delivery flow status (pulsation). If pulsation is large, rinse the flow ch... Show Detail
- No:4299
- Date of Issue:2020/12/03 11:05
- Category: Common , Detector , Solvent Delivery Module
- Brached FAQ
-
Check the light level. If the light level is low even with an air cell or if the lamp has been operated for more than 2000 hours, then replace the lamp. ... Show Detail
- No:4298
- Date of Issue:2020/12/03 11:03
- Category: Detector
- Brached FAQ
-
Specify an appropriate range value. Set [LAMP] to [D2]. The light source lamp illuminates. Click the [ZERO] button in the [RUN] window. Adjust the zero p... Show Detail
- No:4297
- Date of Issue:2020/12/03 11:01
- Category: Detector
- Brached FAQ
-
The baseline undulates with the pump stroke
Click [PURGE] to purge the bubbles. Alternatively, connect a syringe to the drain tube connection port to suction out the bubbles. Click here for Manual P... Show Detail
- No:4296
- Date of Issue:2020/12/03 11:00
- Category: Common , Solvent Delivery Module
- Brached FAQ
-
Check for leakage from cell inlet and outlet pipe connections and from the cell window. If there is leakage, disassemble and clean the cell. If there is lo... Show Detail
- No:4295
- Date of Issue:2020/12/03 10:59
- Category: Common
- Brached FAQ