FAQ - Frequently Asked Questions
- Products Search > GC/GCMS Helium Shortage
Keyword Search
Frequently Referenced FAQ
FAQ within [ GC/GCMS Helium Shortage ]
-
Simulation of gas consumption reduction
You can simulate it on the following pages. [GCMS] https://www.shimadzu.com/an/eco_sim/eco.htm Show Detail
- No:9041
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/27 07:02
-
We are proposing a hydrogen gas generator from Peak Scientific. (* Performance is not guaranteed by Shimadzu Corporation) PEAK Scientific hydrogen ... Show Detail
- No:9042
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:37
-
There are no recommended nitrogen gas generators that have been tested in our company. The nitrogen gas generator recommended by our company for LCMS is n... Show Detail
- No:9043
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:38
-
See below. https://www.shimadzu.com/an/gc/safety_function.html Show Detail
- No:9044
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:39
-
Sensitivity variation by changing carrier gas
[GC] Although the sensitivity depends on the detector, some can achieve comparable sensitivity with helium carriers and hydrogen or nitrogen carriers. Als... Show Detail
- No:9045
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:40
-
Method of alternative carrier gas
Link to EZGC Method Translator (external page) Restek's EZGC Method Translator at the above URL is useful for considering methods when using alternative ... Show Detail
- No:9046
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:41
-
[Hydrogen] Existing mass spectral libraries can be used. However, it has been reported that some compounds may change the ratio of ions. [Nitrogen] ... Show Detail
- No:9047
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:43
-
The following models are compatible with hydrogen sensor. [GC] GC-2030 [GCMS] GCMS-QP2020NX, GCMS-TQ8040NX, GCMS-TQ8050NX Show Detail
- No:9048
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:44
-
About Dedicated Systems/Database Products
Some products/analyses can be followed. Contact your our company sales representative. Show Detail
- No:9049
- Date of Issue:2022/10/20 08:22
-
The Py-screener is not compatible with alternative carrier gases. However, if you have the Auto-Shot Sampler for Py-3030D in your system, it is possible to c... Show Detail
- No:9050
- Date of Issue:2022/10/20 08:22
- Last Updated:2022/10/20 08:44
1-10 of Total 11