FAQ - Frequently Asked Questions
- Products Search > Non-Destructive Testing (NDT) Systems > inspeXio SMX-225CT
Keyword Search
Frequently Referenced FAQ
FAQ within [ inspeXio SMX-225CT ]
-
(SMX-225CT) Rotating the X-ray Tube Head Tungsten Target
This system emits X-rays by irradiating a target with a focused electron beam. The tungsten can melt and vaporize, caused a rough targetsurface, which can... Show Detail
- No:8680
- Date of Issue:2022/05/23 09:11
-
(SMX-225CT) Cleaning the X-ray Tube Vacuum Chamber (Video)
Clean the surface of the focus cup and the inside of the X-ray tube vacuum chamber. Perform this procedure when the filament is replaced or when aging does ... Show Detail
- No:8678
- Date of Issue:2022/05/23 09:04
- Last Updated:2022/05/26 14:23
-
(SMX-225CT) Replacing the Filament
1. Click thebutton on the X-ray control area to display the [X-Ray MaintenanceSchedule] window. 2. Click the [Reset] button in [Usage time after ... Show Detail
- No:8677
- Date of Issue:2022/05/20 14:00
- Last Updated:2022/05/26 14:25
- Brached FAQ
-
(SMX-225CT) Status on the X-ray Panel Unit
The three-digit status display on the X-ray panel unit can display a blinking code, whichindicates an error state, in addition to the solidly indicated 000 a... Show Detail
- No:8674
- Date of Issue:2022/05/19 16:22
-
(SMX-225CT) Aging the X-ray generator
The CT software inspeXio64 determines the tube voltage output permitted based on thelast time the previous aging process was finished and the specified volta... Show Detail
- No:8666
- Date of Issue:2022/05/19 11:39
-
(SMX-225CT) Startup Inspections
1. Inspect the lead glass for any scratches or cracks. If nothing unusual isdiscovered, click the [Next] button. WARNING:If the lead glass is scratc... Show Detail
- No:8665
- Date of Issue:2022/05/19 11:18
-
(SMX-225CT) Restoring the System After an Error Shutdown
1. Close the CT software. (If the software cannot be closed, go to Step 2.) 2. Turn the [OPERATE] key switch counterclockwise to the [OFF] position. ... Show Detail
- No:8661
- Date of Issue:2022/05/19 10:44
- Brached FAQ
-
(SMX-225CT) Shutting Down the System
CAUTION:If the system will be operated intermittently (without long periods of disuse),shut down the system according to the procedure described up to "Shutt... Show Detail
- No:8664
- Date of Issue:2022/05/19 10:43
-
(SMX-225CT) Supplying Power to the System
CAUTION: If operating the system intermittently (without a long period of disuse), startup the system beginning with "Supplying Power Inside the System". I... Show Detail
- No:8662
- Date of Issue:2022/05/19 10:32
-
(SMX-225CT) Starting the Software
1. Checking for Samples on the Stage The CT stage is initialized during CT software inspeXio64 startup. CT stage initialization consists of moving the ... Show Detail
- No:8663
- Date of Issue:2022/05/19 10:22
11-20 of Total 21