  • (MCE) The peak detection position fluctuates

    •The temperature of the separation buffer was not sufficiently at room temperature. →Allow 30 minutes or more to elapse the separation buffer at room temp... Show Detail

    • No:8363
    • Date of Issue:2022/02/16 16:24
    • Last Updated:2022/02/25 16:11
  • (MCE) Peaks do not separate

    •The correct separation buffer is notused. →Check the reagent kit. •The separation buffer or marker hasdeteriorated. →Verify that they have been... Show Detail

    • No:8359
    • Date of Issue:2022/02/16 16:24
    • Last Updated:2022/02/25 16:13
  • (MCE) The baseline fluctuates

    •Separation buffer and dye are not mixed evenly. →Add dye to a new batch of separation buffer, andagitate it sufficiently. •Rinse water is insuffic... Show Detail

    • No:8357
    • Date of Issue:2022/02/16 16:24
  • (MCE) Emergency Measures / During a Power Outage

    « Emergency Measures» If a strange noise or smell emanates from the instrument, or some other irregularity is noticed, stopinstrument operation using the ... Show Detail

    • No:8318
    • Date of Issue:2022/02/16 15:35
  • (MCE) Instrument Storage

    1.Rinse Water Removal Follow the procedures below to remove rinse water in the rinse water suction line prior to moving or storingthe instrument. 1) Re... Show Detail

    • No:8354
    • Date of Issue:2022/02/16 13:46
  • (MCE) Parts Maintenance and Replacement

    « Microchips» Select [Chip Management] from the [Instrument] menu in the MultiNA window. The microchip usage status (number of runs, date and time of fir... Show Detail

    • No:8351
    • Date of Issue:2022/02/16 13:20
  • (MCE) Replacement and Maintenance of Suction Filter

    « Guidelines for Procedure» • When a significant number of bubbles are present in the rinse water suction line. • When the suction filter appears to be dir... Show Detail

    • No:8349
    • Date of Issue:2022/02/16 11:56
  • (MCE) Replacement and Inspection of Pump Cartridge

    « Requirements» • Pump cartridge • Phillips screwdriver • Wipes « Replacement Procedure» Replacement of cartridge No. 3 is used as the exam... Show Detail

    • No:8347
    • Date of Issue:2022/02/16 08:42
    • Brached FAQ
  • (MCE) Replacement and Inspection of Syringe

    « Requirements» • Plunger: Part Number228-25237-14 • Wipes • Beaker • Milli-Q ultrapure water • Phillips screwdriver « Replacement Procedu... Show Detail

    • No:8345
    • Date of Issue:2022/02/16 08:42
    • Brached FAQ
  • (MCE) Replacement and Inspection of Plunger

    « Requirements» • Plunger: Part Number228-25237-14 • Wipes • Beaker • Milli-Q ultrapure water • Phillips screwdriver « Replacement Procedu... Show Detail

    • No:8344
    • Date of Issue:2022/02/16 08:42
    • Brached FAQ