• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ MCE-202(MultiNA) ]

11-20 of Total 31

≪ 2 / 4page ≫
  • (MCE) Chip Cleaning

    1. Dispense chip cleaning solution into the buffer bottle. NOTE: The required volume of chip cleaning solution varies depending on conditions such as the n... Show Detail

    • No:8341
    • Date of Issue:2022/02/15 15:00
    • Brached FAQ
  • (MCE) Replacement and Inspection of Plunger

    « Requirements» • Plunger: Part Number228-25237-14 • Wipes • Beaker • Milli-Q ultrapure water • Phillips screwdriver « Replacement Procedu... Show Detail

    • No:8344
    • Date of Issue:2022/02/16 08:42
    • Brached FAQ
  • (MCE) Replacement and Inspection of Syringe

    « Requirements» • Plunger: Part Number228-25237-14 • Wipes • Beaker • Milli-Q ultrapure water • Phillips screwdriver « Replacement Procedu... Show Detail

    • No:8345
    • Date of Issue:2022/02/16 08:42
    • Brached FAQ
  • (MCE) Replacement and Inspection of Pump Cartridge

    « Requirements» • Pump cartridge • Phillips screwdriver • Wipes « Replacement Procedure» Replacement of cartridge No. 3 is used as the exam... Show Detail

    • No:8347
    • Date of Issue:2022/02/16 08:42
    • Brached FAQ
  • (MCE) Replacement and Maintenance of Suction Filter

    « Guidelines for Procedure» • When a significant number of bubbles are present in the rinse water suction line. • When the suction filter appears to be dir... Show Detail

    • No:8349
    • Date of Issue:2022/02/16 11:56
  • (MCE) Parts Maintenance and Replacement

    « Microchips» Select [Chip Management] from the [Instrument] menu in the MultiNA window. The microchip usage status (number of runs, date and time of fir... Show Detail

    • No:8351
    • Date of Issue:2022/02/16 13:20
  • (MCE) Instrument Storage

    1.Rinse Water Removal Follow the procedures below to remove rinse water in the rinse water suction line prior to moving or storingthe instrument. 1) Re... Show Detail

    • No:8354
    • Date of Issue:2022/02/16 13:46
  • (MCE) Emergency Measures / During a Power Outage

    « Emergency Measures» If a strange noise or smell emanates from the instrument, or some other irregularity is noticed, stopinstrument operation using the ... Show Detail

    • No:8318
    • Date of Issue:2022/02/16 15:35
  • (MCE) The baseline fluctuates

    •Separation buffer and dye are not mixed evenly. →Add dye to a new batch of separation buffer, andagitate it sufficiently. •Rinse water is insuffic... Show Detail

    • No:8357
    • Date of Issue:2022/02/16 16:24
  • (MCE) Peaks do not separate

    •The correct separation buffer is notused. →Check the reagent kit. •The separation buffer or marker hasdeteriorated. →Verify that they have been... Show Detail

    • No:8359
    • Date of Issue:2022/02/16 16:24
    • Last Updated:2022/02/25 16:13

11-20 of Total 31

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.