1. Dispense chip cleaning solution into the buffer bottle. NOTE: The required volume of chip cleaning solution varies depending on conditions such as the n... Show Detail
- No:8341
- Date of Issue:2022/02/15 15:00
- Brached FAQ
-
(MCE) Replacement and Inspection of Plunger
« Requirements» • Plunger: Part Number228-25237-14 • Wipes • Beaker • Milli-Q ultrapure water • Phillips screwdriver « Replacement Procedu... Show Detail
- No:8344
- Date of Issue:2022/02/16 08:42
- Brached FAQ
-
(MCE) Replacement and Inspection of Syringe
« Requirements» • Plunger: Part Number228-25237-14 • Wipes • Beaker • Milli-Q ultrapure water • Phillips screwdriver « Replacement Procedu... Show Detail
- No:8345
- Date of Issue:2022/02/16 08:42
- Brached FAQ
-
(MCE) Replacement and Inspection of Pump Cartridge
« Requirements» • Pump cartridge • Phillips screwdriver • Wipes « Replacement Procedure» Replacement of cartridge No. 3 is used as the exam... Show Detail
- No:8347
- Date of Issue:2022/02/16 08:42
- Brached FAQ
-
(MCE) Replacement and Maintenance of Suction Filter
« Guidelines for Procedure» • When a significant number of bubbles are present in the rinse water suction line. • When the suction filter appears to be dir... Show Detail
- No:8349
- Date of Issue:2022/02/16 11:56
-
(MCE) Parts Maintenance and Replacement
« Microchips» Select [Chip Management] from the [Instrument] menu in the MultiNA window. The microchip usage status (number of runs, date and time of fir... Show Detail
- No:8351
- Date of Issue:2022/02/16 13:20
-
1.Rinse Water Removal Follow the procedures below to remove rinse water in the rinse water suction line prior to moving or storingthe instrument. 1) Re... Show Detail
- No:8354
- Date of Issue:2022/02/16 13:46
-
(MCE) Emergency Measures / During a Power Outage
« Emergency Measures» If a strange noise or smell emanates from the instrument, or some other irregularity is noticed, stopinstrument operation using the ... Show Detail
- No:8318
- Date of Issue:2022/02/16 15:35
-
•Separation buffer and dye are not mixed evenly. →Add dye to a new batch of separation buffer, andagitate it sufficiently. •Rinse water is insuffic... Show Detail
- No:8357
- Date of Issue:2022/02/16 16:24
-
•The correct separation buffer is notused. →Check the reagent kit. •The separation buffer or marker hasdeteriorated. →Verify that they have been... Show Detail
- No:8359
- Date of Issue:2022/02/16 16:24
- Last Updated:2022/02/25 16:13