(LCMS) Changing the Power Interruption Automatic Recovery...
This content is for LCMS-2050. 1. Click [System Configuration] on the [Main] assistant bar. 2. Double-click “LCMS: LCMS-2050” displayed in the [Modules... Show Detail
- No:7908
- Date of Issue:2022/01/24 16:22
-
(LCMS) Changing the Settings for the Gas and Pump Interlo...
This content is not applied to LCMS-9030. Protective functionality is available for preventing MS contamination by automatically switching OFF the LC pump... Show Detail
- No:7910
- Date of Issue:2022/01/25 08:30
-
(LCMS) Changing Settings for the Detector Protection Func...
This content is not applied to LCMS-9030. Protective functionality is available for automatically switching OFF the detector if too many ions enter the de... Show Detail
- No:7912
- Date of Issue:2022/01/25 08:30
-
(LCMS) Copying Data Files Automatically
Data files (.lcd) can be copied to a specified folder during data acquisition or when files are updated. On the [Tools] menu in the [Data Acquisition] window... Show Detail
- No:7916
- Date of Issue:2022/01/25 08:30
-
(LCMS) LC Pump Switches OFF on its Own
The LC pump cannot deliver solvent if the MS unit is not ON. Switch the pump ON after starting up the MS unit. However, if the gas and pump inter... Show Detail
- No:7914
- Date of Issue:2022/01/25 08:43
-
(LCMS) Displaying MS Chromatograms
Right-click on the [Chromatogram View] in the [MS Data Analysis] window and then click [Data View Parameters] on the menu (or on the toolbar). Click the [... Show Detail
- No:7918
- Date of Issue:2022/01/25 08:53
-
In the [MS Data Analysis] window, click the [Spectrum View]-[Display] tab, and then select the checkboxes for the event numbers to display. To overlay spect... Show Detail
- No:7920
- Date of Issue:2022/01/25 09:02
-
(LCMS) Relationship between Method, Batch, and Data Files
Method files (.lcm, .gcm) contain data acquisition conditions, analysis conditions, calibration curve settings, and other information. Batch files (.lcb, .g... Show Detail
- No:7922
- Date of Issue:2022/01/25 09:05
-
(LCMS) Checking the MS Instrument Status
This content is for LCMS-2050. The [System Check]-[MS Check] function can automatically check or determine the status of MS detectors (mass axis shift, FW... Show Detail
- No:7927
- Date of Issue:2022/01/25 10:46
- Last Updated:2022/06/24 12:50
-
(LCMS) Changing the System Check Pass/Fail Criteria
System check criteria can be changed by following the procedure below. Open the [System Configuration] sub-window, select the LCMS system in the [Modules ... Show Detail
- No:7928
- Date of Issue:2022/01/25 10:49