• Font Size
  • S
  • M
  • L

FAQ within [ EZ-Test Series ]

21-27 of Total 27

3 / 3page
  • (EZ-Test) Some buttons don't work

    • The types of buttons that can be operated is restricted during tests or whileconnected to a computer. Try again after the test is finished or afterdisconne... Show Detail

    • No:7685
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:32
  • (EZ-Test) A servo error occurred during jog operations

    • The motor may be overloaded. Confirm there are no obstructions caughtbetween the crosshead and table. Also confirm that nothing is wrapped aroundthe ball s... Show Detail

    • No:7684
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:32
  • (EZ-Test) A servo error occurred during testing

    • The motormay be overloaded. Confirm whether the testing machine operatinglimits have been exceeded. (Max 10 hours continuous operation, max 1000cycles, or ... Show Detail

    • No:7683
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:31
  • (EZ-Test) Motor does not turn

    • Confirm whether the emergency stop button is ON. Click here for "(EZ-Test) Safety Systems". • Confirm whether a stroke limit was triggered. The moto... Show Detail

    • No:7682
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:31
  • (EZ-Test) A ground fault circuit interrupter was tripped ...

    • There may be an electrical leak within theEZ-X unit. Immediately stop usingthe system and contact a Shimadzu sales or service representative. Show Detail

    • No:7681
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:30
  • (EZ-Test) Fuses fail frequently

    • Confirm whether the input voltage mode switch is set properly. • Confirm whether the fuses current rating is correct.(100V:10A, 220V:5A) • Confirm whethe... Show Detail

    • No:7680
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:30
  • (EZ-Test) Cannot switch the instrument power ON

    • Confirm whether a circuit breaker has tripped on the power supply panel. • Confirm whether the power cordis disconnected. • Confirm whether the input vol... Show Detail

    • No:7679
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:29

21-27 of Total 27

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.