  • (EZ-Test) Alarm Log

    The main frame stores the alarms have occurred in the last 20 times. The alarm log displays in order fromthe latest. • Pressing the jog button allows... Show Detail

    • No:7691
    • Date of Issue:2022/01/20 09:08
  • (EZ-Test) Alarm Display / Alarm Code

    If an alarm occurs in theEZ-X unit , an alarm sound i s emitted and the alarm screen flashes. Tocancel the alarm, eliminate the cause and press the [STOP]... Show Detail

    • No:7676
    • Date of Issue:2022/01/19 14:58
    • Last Updated:2022/01/21 12:40
  • (EZ-Test) Safety Systems

    Once upper and lower mechanical crosshead limits are set, this feature force stops the crossheadby shutting OFF th e motor drive power if the crosshead excee... Show Detail

    • No:7674
    • Date of Issue:2022/01/19 14:39
    • Brached FAQ
  • (EZ-Test) Attaching Load Cells

    1. Switch the EZ-X power OFF. 2. Remove any jigs attached to the load cell. 3. Loosen the two fastener screws on the CAL ca ble conn ector on the right sid... Show Detail

    • No:7669
    • Date of Issue:2022/01/19 13:22
    • Brached FAQ
  • (EZ-Test) Calibrating Test Force (E-CAL)

    Electronically calibrates (E-CAL) the test force. Calibrating the test force with E-CAL usesprecision resistors built into the calibration cable to apply a s... Show Detail

    • No:7668
    • Date of Issue:2022/01/19 13:10
    • Last Updated:2022/01/19 13:41
  • (EZ-Test) In an Emergency / During a Power Outage

    If a malfunction occurs, such as the crosshead not stopping, a burning odor being emitted, ora power failure, take the following actions. Then inspect the s... Show Detail

    • No:7665
    • Date of Issue:2022/01/19 11:36
    • Last Updated:2022/01/21 10:40
  • (EZ-Test) Resetting the System

    Resetting restores allEZ-X settings to factory settings. NOTE: • After restoring factory settings, optional devices must be registered, and the sys... Show Detail

    • No:7689
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 11:10
  • (EZ-Test) An optional device is not functioning

    •Confirm whether the presence of the optional device was registered correctly.Also, some optional devices require restarting the testing machine. Switch theE... Show Detail

    • No:7688
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:38
  • (EZ-Test) The test force display indicates "---"

    • The load cell is not being recognized. Temporarily switch the power OFF andconfirm that the calibration cable is connected properly, then switch the powerO... Show Detail

    • No:7687
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:33
  • (EZ-Test) None of the buttons work

    • No buttons function while theEZ-X unit is starting up. Wait until it finishesstartup. Show Detail

    • No:7686
    • Date of Issue:2020/12/02 14:16
    • Last Updated:2022/01/26 10:32