FAQ within [ AGS-X Series ]
-
(AGS-X) Attaching Compression and Bending Test Jigs
1. Screw in the compression loading jig. 2.Temporarily loosen the set bolt, theninsert the upper compression plate. 3.Insert the boss on the lowercompr... Show Detail
- No:7999
- Date of Issue:2022/01/20 11:26
- Brached FAQ
-
1. Attaching the Specimen 1) Insert the specimen at thecenter of the grip faces. 2) Tighten securely with bothhands. 3) Use jog operations to... Show Detail
- No:7997
- Date of Issue:2022/01/20 10:55
-
(AGS-X) Attaching a Tensile Test Jig
« Attaching the Upper grip (Non-shift Wedge Type Grips, Screw Type Flat Grips» 1. Screw the tensile joint intothe buttom surface of theload cell. 2.Inser... Show Detail
- No:7994
- Date of Issue:2022/01/20 09:24
- Brached FAQ
-
The main frame stores the alarms have occurred in the last 20 times. The alarm log displays in order fromthe latest. • Pressing the jog button allows... Show Detail
- No:7941
- Date of Issue:2022/01/20 09:08
-
(AGS-X) Alarm Display / Alarm Code
If an alarm occurs in the AGS-X unit , an alarm sound i s emitted and the alarm screen flashes. Tocancel the alarm, eliminate the cause and press the [STO... Show Detail
- No:7943
- Date of Issue:2022/01/19 14:58
-
(AGS-X) Calibrating Test Force (E-CAL)
Electronically calibrates (E-CAL) the test force. Calibrating the test force with E-CAL usesprecision resistors built into the calibration cable to apply a s... Show Detail
- No:7936
- Date of Issue:2022/01/19 13:10
-
Resetting restores all AGS-Xsettings to factory settings. NOTE: • After restoring factory settings, optional devices must be registered, and the sy... Show Detail
- No:7991
- Date of Issue:2020/12/02 14:16
-
(AGS-X) An optional device is not functioning
•Confirm whether the presence of the optional device was registered correctly.Also, some optional devices require restarting the testing machine. Switch the ... Show Detail
- No:7989
- Date of Issue:2020/12/02 14:16
-
(AGS-X) The test force display indicates "---"
• The load cell is not being recognized. Temporarily switch the power OFF andconfirm that the calibration cable is connected properly, then switch the powerO... Show Detail
- No:7987
- Date of Issue:2020/12/02 14:16
-
(AGS-X) None of the buttons work
• No buttons function while the AGS-Xunit is starting up. Wait until it finishesstartup. Show Detail
- No:7985
- Date of Issue:2020/12/02 14:16