FAQ - Frequently Asked Questions
- Products Search > Materials Testing & Inspection > Autograph(AG) > AGS-X Series
Keyword Search
Frequently Referenced FAQ
FAQ within [ AGS-X Series ]
-
(AGS-X) Cannot switch the instrument power ON
• Confirm whether a circuit breaker has tripped on the power supply panel. • Confirm whether the power cordis disconnected. • Confirm whether the input vol... Show Detail
- No:7947
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:28
-
• Confirm whether the input voltage mode switch is set properly. • Confirm whether the fuses current rating is correct.(100V/120V:10 A, 220V/240V:5A) • Con... Show Detail
- No:7972
- Date of Issue:2020/12/02 14:16
- Last Updated:2022/01/26 10:48
-
(AGS-X) A ground fault circuit interrupter was tripped on...
• There may be an electrical leak within the AGS-Xunit. Immediately stop usingthe system and contact a Shimadzu sales or service representative. Show Detail
- No:7975
- Date of Issue:2020/12/02 14:16
-
• Confirm whether the emergency stop button is ON. Click here for "(AGS-X) Safety Systems". • Confirm whether a stroke limit was triggered. The motor ... Show Detail
- No:7977
- Date of Issue:2020/12/02 14:16
-
(AGS-X) A servo error occurred during testing
• The motormay be overloaded. Confirm whether the testing machine operatinglimits have been exceeded. (Max 10 hours continuous operation, max 1000cycles, or ... Show Detail
- No:7979
- Date of Issue:2020/12/02 14:16
-
(AGS-X) A servo error occurred during jog operations
• The motor may be overloaded. Confirm there are no obstructions caughtbetween the crosshead and table. Also confirm that nothing is wrapped aroundthe ball s... Show Detail
- No:7981
- Date of Issue:2020/12/02 14:16
-
(AGS-X) Some buttons don't work
• The types of buttons that can be operated is restricted during tests or whileconnected to a computer. Try again after the test is finished or afterdisconne... Show Detail
- No:7983
- Date of Issue:2020/12/02 14:16
-
(AGS-X) None of the buttons work
• No buttons function while the AGS-Xunit is starting up. Wait until it finishesstartup. Show Detail
- No:7985
- Date of Issue:2020/12/02 14:16
-
(AGS-X) The test force display indicates "---"
• The load cell is not being recognized. Temporarily switch the power OFF andconfirm that the calibration cable is connected properly, then switch the powerO... Show Detail
- No:7987
- Date of Issue:2020/12/02 14:16
-
(AGS-X) An optional device is not functioning
•Confirm whether the presence of the optional device was registered correctly.Also, some optional devices require restarting the testing machine. Switch the ... Show Detail
- No:7989
- Date of Issue:2020/12/02 14:16
1-10 of Total 27