FAQ within [ AGX-V Series ]
-
You can return the instrument settings to the factory default when you need to resetthe system parameters to the initial values collectively or when incorrec... Show Detail
- No:8070
- Date of Issue:2020/12/02 14:16
-
(AGX-V) The instrument does not turn on
• Check that the breaker of the distribution board is on. • Check that the power supply cable is not loose. Show Detail
- No:8073
- Date of Issue:2020/12/02 14:16
-
(AGX-V) The leakage breaker of the instrument is activated
There may be leakage inside the instrument. Immediatelystop using the instrument and contact a Shimadzu sales or service representative. Show Detail
- No:8075
- Date of Issue:2020/12/02 14:16
-
(AGX-V) The leakage breaker of the distribution board is ...
There may be leakage inside the instrument. Immediatelystop using the instrument andcontact a Shimadzu sales or service representative. Show Detail
- No:8077
- Date of Issue:2020/12/02 14:16
-
(AGX-V) The motor does not operate
• Check that the emergency stop switch is not active. • Check that the crosshead limit switches are not active. • The motor does not operate while the inst... Show Detail
- No:8079
- Date of Issue:2020/12/02 14:16
-
(AGX-V) A servo error occurs during a test
Overload may have occurred in the motor. Check whetherthe instrument is used within the working limits. (Max. tenhours continuous operation, max. 1,000 re... Show Detail
- No:8081
- Date of Issue:2020/12/02 14:16
-
(AGX-V) A servo error occurs during movement in the jog o...
• Overload may have occurred in the motor. • Check that noforeign matter is caught between the crosshead and table. • Check that no foreign matter is win... Show Detail
- No:8083
- Date of Issue:2020/12/02 14:16
-
(AGX-V) Some keys do not accept input
Active keys are limited during a test or when a computeris connected. Try again after a test ends or when acomputer is not connected. When a computer is conn... Show Detail
- No:8085
- Date of Issue:2020/12/02 14:16
-
(AGX-V) No key input is accepted
No key input is accepted while the instrument is startingup. Wait until the startup is completed. Show Detail
- No:8087
- Date of Issue:2020/12/02 14:16
-
(AGX-V) The test force display is "- - - -"
• The load cell is not recognized. Turn off the power, checkthat the calibration cable is connected appropriately, andturn on the power again. • If the type... Show Detail
- No:8089
- Date of Issue:2020/12/02 14:16