FAQ - Frequently Asked Questions
- Products Search > Materials Testing & Inspection > Universal Testing Machine(UH)
Keyword Search
Frequently Referenced FAQ
FAQ within [ Universal Testing Machine(UH) ]
-
Alarm Codes If the testing machine detects an abnormality, it displays the Alarm screen. While an alarm massage is displayed, the testing machine stops... Show Detail
- No:7547
- Date of Issue:2021/12/03 10:07
-
(UH) Maintenance of the Hydraulic Unit
This is a high-pressure fixed-discharge type hydraulic unit incorporating a high-pressure type gear pump. Itmainly consists of a hydraulic pump, a pump motor... Show Detail
- No:7546
- Date of Issue:2021/12/03 09:02
-
In this machine, the past 50 event logs that are generated are saved. Event logs are subdivided intothe following classes. • Behavior of the testing machine... Show Detail
- No:7543
- Date of Issue:2021/12/03 09:02
-
As for Items as follows, you can check the test machine as to whether they have troubles. If you think that they might have troubles, please use this functio... Show Detail
- No:7542
- Date of Issue:2021/12/03 09:02
-
(UH) Checking of the emergency stop equipment
Before using this machine, please check the emergency stop equipment. If this machine didn’t work in the undermentioned check, contact a Shimadzu sales or se... Show Detail
- No:7541
- Date of Issue:2021/12/03 09:02
-
(UH) Adjusting the Zero Point of the Test Force (Auto Zero)
The zero adjustment of the test force is executed automatically by a touch-panel operation (Auto Zero function). CAUTION: • The zero adjustment of... Show Detail
- No:7540
- Date of Issue:2021/12/02 16:22
-
(UH) Electronic Calibration (E-CAL)
The electronic calibration (E-CAL) is required once per day before the daily startup of the testingmachine. Even so, this calibration must be done immediatel... Show Detail
- No:7539
- Date of Issue:2021/12/02 16:22
-
(UH) Mounting the platens for Compression Tests
CAUTION:When attaching the compression platen, be carefull not to have your fingerscaught. Be carefull not to fall the compression platen. Compression pl... Show Detail
- No:7538
- Date of Issue:2021/12/02 16:09
- Brached FAQ
-
(UH) Using the Grip jaws for Tensile Tests
CAUTION:The grip jaws have sharp edges to allow grasping of the specimen. When handling the grip jaws, be careful not to cut your fingers or hand. ... Show Detail
- No:7537
- Date of Issue:2021/12/02 15:53
- Brached FAQ
-
1. Levelness inspection of the load frame 1)Place a spirit level near the center of the table top and measure the levelness of the table in all dire... Show Detail
- No:7545
- Date of Issue:2021/12/02 15:17
- Brached FAQ
1-10 of Total 16