  • (DUH) In an Emergency

    If you find an abnormal condition such as smoke or you smell something burning, take the following measures: To restart operation, check the equipment, an... Show Detail

    • No:7513
    • Date of Issue:2021/12/06 09:44
  • (DUH) Illumination Lamp Replacement

    Part Number:347-24149-01 Illumination Lamp 1. Remove upper cover. 2.Remove lamp ASSY and install new lamp ASSY. 3.Close the top cover, an... Show Detail

    • No:7514
    • Date of Issue:2021/12/06 10:05
  • (DUH) Indenter Replacement

    When the tip of the indenter breaks or you wish to replace the indenter with a different type, perform the following procedure. Part Number:340-47013 Triang... Show Detail

    • No:7515
    • Date of Issue:2021/12/06 10:05
  • (DUH) Centering an Indentation

    If the indenter lowering position is deviated from the center of the visual field, perform the following procedure: 1. Remove the front cover. 1) R... Show Detail

    • No:7516
    • Date of Issue:2021/12/06 10:05
  • (DUH) Indenter Height Adjustment

    When the microscope focal point is adjusted and the stage is moved to under the indenter before a test is started, if the indenter touches the sample or the ... Show Detail

    • No:7518
    • Date of Issue:2021/12/06 10:05
  • (DUH) Objective Lens Replacement

    Replacing the x50 objective lens: Part Number:344-89964-40 Objective lensx50 If the objective lens is damaged or the lens visibility is deteriorated ... Show Detail

    • No:7520
    • Date of Issue:2021/12/06 10:05
  • (DUH) Identical Focal Point Adjustment

    1. While observing a sample through the x50 objective lens, move the stage up/down by using the elevating handle to perform focal point adjustment. 2. Switc... Show Detail

    • No:7521
    • Date of Issue:2021/12/06 10:05
  • (DUH) Magnification Adjustment

    1. Make sure that the cables are properly connected. Turn ON the power supply. And please let me turn on a lamp. When you do not switch on the light, please ... Show Detail

    • No:7522
    • Date of Issue:2021/12/06 11:21
    • Brached FAQ
  • (DUH) Indentation shape is abnormal

    If the Indentation shape is abnormal as shown in below, the indenter mounting part may be defective. Replace the defective part with a new one. For replaceme... Show Detail

    • No:7523
    • Date of Issue:2021/12/06 11:48
  • (DUH) Trouble during the test

    1. “Slide sample stage to load side” is displayed at the start of test • Causes1:The stage has not been moved to under the indenter. Before starting ... Show Detail

    • No:7524
    • Date of Issue:2021/12/06 11:48