• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ Micro Hardness Tester(HMV) ]

21-25 of Total 25

≪ 3 / 3page ≫
  • (HMV) Adjusting the Diamond Indenter Attachment

    Part Number:347-20344VICKERS INDENTOR 100D Part Number:340-06619-07STD TEST BLOCK HMV-700 1. Problems with the Shape of the Indentation If the sh... Show Detail

    • No:7475
    • Date of Issue:2021/12/06 10:05
  • (HMV) Replacing the Illumination Lamp

    PartNumber:347-24149-02LED LAMP ASSY 1. Remove the top cover fastening screws, and remove the top cover. 2. Remove the fastening screws. Remove the l... Show Detail

    • No:7474
    • Date of Issue:2021/12/06 10:05
  • (HMV) Troubleshooting (HMV-G-MM)

    Problem1:The program does notstart up. Cause1:Part of the program was deleted, orthe folder was changed. Solution1:Reinstall the program. Pr... Show Detail

    • No:7473
    • Date of Issue:2021/12/06 09:44
  • (HMV) Maintenance and Inspection (HMV-G-MM)

    • Maintenance 1. Turn OFF the power to the hardness tester and the motorized micrometer kit controllerBOX. 2. Turning the knob of the motorized micrometer ... Show Detail

    • No:7472
    • Date of Issue:2021/12/06 09:44
  • (HMV) In an Emergency

    If you notice smoke, a burnt smell, or anything else unusual, take the following steps. Before using the instrument again, inspect it and, if necessary, con... Show Detail

    • No:7471
    • Date of Issue:2021/12/06 09:44

21-25 of Total 25

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.