FAQ - Frequently Asked Questions
- Products Search > Materials Testing & Inspection > Hardness Tester > Micro Hardness Tester(HMV)
Keyword Search
Frequently Referenced FAQ
FAQ within [ Micro Hardness Tester(HMV) ]
-
(HMV) If the illumination lamp does not light
Check the lamp as "Replacing the Illumination Lamp". If it is burnt out, replace it. If despite this it still will not light, contact a Shimadzu sales or s... Show Detail
- No:7501
- Date of Issue:2021/12/06 10:05
-
(HMV) If the standard value of a hardness test piece does...
Possible Causes: 1. Effects of vibrations Vibrations will have a significant effect at test forces of 98.07 mN or less. Ensure that thearea around the i... Show Detail
- No:7502
- Date of Issue:2021/12/06 10:05
-
(HMV) If the test force or retention time is different th...
Check the test conditions. The effect of noise can result in settings with abnormal values. Ifthe test condition values are as expected but the values dur... Show Detail
- No:7503
- Date of Issue:2021/12/06 10:05
-
(HMV) If the indentation readings are abnormal
Adjust the magnification as per section "Replacing and Adjusting the Objective Lens". Click here for "(HMV) Replacing and Adjusting the Objective Lens".... Show Detail
- No:7504
- Date of Issue:2021/12/06 10:05
-
(HMV) If the clock time is incorrect
1.The room temperature has dropped below freezing. This is not a malfunction. The clock IC does not operate at temperatures below freezing. 2.The ... Show Detail
- No:7505
- Date of Issue:2021/12/06 10:05
-
(HMV) The ready lamp and the running lamp blink
Replace the battery on the CPU board. Contact a Shimadzu sales or service representative. Show Detail
- No:7506
- Date of Issue:2021/12/06 10:05
-
(HMV) Replacing and Adjusting the Objective Lens
Adjust the magnification as indicated below if the optical head has been removed, if theobjective lens has been replaced, or if the magnification is unsuitab... Show Detail
- No:7478
- Date of Issue:2021/12/06 11:21
- Brached FAQ
-
(HMV) Camera Attachment Position and Focusing Adjustments
Before performing magnification and image center adjustments, adjust the camera's attachment position until the image observed through the tester's eyepiece ... Show Detail
- No:7481
- Date of Issue:2021/12/06 11:48
-
This adjustment determines the position of the cross-hair displayed in the image window during testing. In Automatic Detection mode, the search for indentati... Show Detail
- No:7482
- Date of Issue:2021/12/06 11:48
-
(HMV) Adjusting the Brightness and Contrast
This adjusts the quality of the image imported to the PC. Perform this adjustment in the following cases: • The tester's illumination lamp has been replace... Show Detail
- No:7485
- Date of Issue:2021/12/06 11:48
11-20 of Total 25