FAQ - Frequently Asked Questions
- Products Search > Particle Size Analysis
Keyword Search
Frequently Referenced FAQ
FAQ within [ Particle Size Analysis ]
-
(SALD) Photometric data is unstable
Cause1:Faulty ground connection Corrective Action1: Use an attached power cord. And, check whether it conforms to the conditions of the installation site.... Show Detail
- No:7468
- Date of Issue:2021/11/18 13:27
-
Cause1:Faulty cell assembly Corrective Action1: The flow cell may not be sealed completely.Reassemble a flow cell. Click here for "(SALD) Sampler, Clean... Show Detail
- No:7469
- Date of Issue:2021/11/18 13:27
-
(SALD) Optical Axis Cannot be Adjusted
Cause1:XY-axisadjustment screw remains locked Corrective Action1:Unlock the XY-axis adjustment screw before making adjustments. Cause2:The XY-axis ... Show Detail
- No:7470
- Date of Issue:2021/11/18 13:27
-
(SALD) Confirmation of Cell Cleaning
Any soiling of the cell may be detected using [Diagnosis & Adjustment]. When this is selected the screen display. Check that the top left of the graph reads ... Show Detail
- No:7444
- Date of Issue:2021/11/18 16:17
-
Before starting to use the analyzer, a hardware check function checks each analyzer function to make sure they all function properly. The hardware checks pr... Show Detail
- No:7445
- Date of Issue:2021/11/19 09:58
-
(SALD) Function Checking Method
The performance check of hardware can be performed on the "Diag & Adjust Screen". If there is problem, contact a Shimadzu sales or service representative.... Show Detail
- No:7446
- Date of Issue:2021/11/19 10:12
-
On the "Diag & Adjust" screen, the following 4 models that can control by PC can do the operation check. • Sampler for SALD-7100: SALD-MS71 • Sampler for S... Show Detail
- No:7447
- Date of Issue:2021/11/19 10:22
-
(SALD) Batch Cell Install/Operation Method
1. Turn off the power switch of SALD. 2. Install the “hole cover” (without hole) to SALD. 3. Open the door of measuring room, pull off the c... Show Detail
- No:7448
- Date of Issue:2021/11/19 11:10
-
(SALD) Batch Cell, Cleaning the batch cell
The daily maintenance tasks required for the device is cleaning the batch cell. Use the cleaning blush (accessory of SALD, Parts number: 046-00906-82 SWAB,D... Show Detail
- No:7449
- Date of Issue:2021/11/19 11:10
-
(SALD) Batch Cell, Troubleshooting
• Problem1:Stir does not move. Cause1-1:Defect connection ofthe power cord Corrective Action1-1: Check if the power cord is correctlyconnected. ... Show Detail
- No:7450
- Date of Issue:2021/11/19 11:10
21-30 of Total 31