(SALD) High-Concentration Measurement Unit, Sample Creati...
1.Sample Creation Procedure 1)Put the sample liquid between the slide glasses. CAUTION:Never use glass cells with cracks or breakage.It may... Show Detail
- No:7456
- Date of Issue:2021/11/18 13:27
-
(SALD) High-Concentration Measurement Unit, Cleaning the ...
Wipe the slide glass softly by the lint-free wiping cloth for industrial use when the outside of slide glass is dirty. CAUTION:Never use glass cells w... Show Detail
- No:7457
- Date of Issue:2021/11/18 13:27
- Last Updated:2021/12/20 11:54
-
(SALD) High-Concentration Measurement Unit, Troubleshooting
Problem: Photometric data shows abnormalities Cause1:セル内に気泡が付着している。 Corrective Action1:Make sample without bubbles. Click here for "(SALD) High... Show Detail
- No:7460
- Date of Issue:2021/11/18 13:27
-
(SALD) Dry measurement case(SALD-DC23) Install/Operation
1. Connect the sensor code to the connector on the wall of measurement room. 2. Remove the sensor unit from SALD-2300. 3. Remove th... Show Detail
- No:7461
- Date of Issue:2021/11/18 13:27
-
(SALD) Dry measurement case(SALD-DC23), Cleaning
Remove the dust tray and clean it. NOTE: Do not clean by alchole, orgenic solvent, acid or alkaline liquid. It causes corrosion. Show Detail
- No:7462
- Date of Issue:2021/11/18 13:27
-
(SALD) Dry measurement case(SALD-DC23), Replace the Filte...
PartNumber: 040-82074-10Filter element 1. Replacement of filter element. 2. Remove the cap of the filter, and replace the filter element... Show Detail
- No:7463
- Date of Issue:2021/11/18 13:27
-
(SALD) Cyclone Injection Type Dry Measurement Unit(SALD-D...
1.Removing condensate in the pressure regulator The air supplied by a compressor contains moisture, dust and oil. The pressure regulator contains a pre... Show Detail
- No:7464
- Date of Issue:2021/11/18 13:27
-
(SALD) Cyclone Injection Type Dry Measurement Unit(SALD-D...
Parts Number: 347-62211Reference sample SALD-2300dry 347-62657-10Sample Cup10mm (3pc set) 347-62657-20Sample Cup 20mm (3pc set) 347-62657-50Sam... Show Detail
- No:7465
- Date of Issue:2021/11/18 13:27
-
(SALD) Cyclone Injection Type Dry Measurement Unit(SALD-D...
Trouble1:Abnormality in measured light intensity data Cause1:Sample is adhered to lens and/or filters Counter measures1:Clean the measuring chamber ... Show Detail
- No:7466
- Date of Issue:2021/11/18 13:27
-
(SALD) Photometric data shows abnormalities
Cause1:Bubbles adhering tothe cell Corrective Action1: Clean the cell using the same type of solvent. Cause2:Defect adjustment of a photo sensor ... Show Detail
- No:7467
- Date of Issue:2021/11/18 13:27